17. If the cDAQ chassis is not reserved automatically, select the chassis and click the Reserve
Chassis button. Refer to the Reserving the Chassis in MAX section for more information.
18. Self-test your chassis in MAX by expanding Devices and Interfaces, right-clicking
NI cDAQ-<model number>, and selecting Self-Test. Self-test performs a brief test to
determine successful chassis installation. When the self-test finishes, a message indicates
successful verification or if an error occurred. If an error occurs, refer to
ni.com/
support/daqmx
.
19. Run a Test Panel in MAX by expanding Devices and Interfaces» NI cDAQ-<model
number>, right-clicking your C Series module, and selecting Test Panels to open a test
panel for the selected module.
If the test panel displays an error message, refer to
ni.com/support.
Click Close to exit the test panel.
Note When in use, the cDAQ chassis may become warm to the touch. This is
normal.
Troubleshooting Chassis Connectivity
If your cDAQ chassis becomes disconnected from the network, try the following:
• After moving the chassis to a new network, NI-DAQmx may lose connection to the chassis.
In this case, click Reconnect to provide NI-DAQmx with the new hostname or IP address.
• The cDAQ chassis icon indicates whether it is recognized and present on the network. If a
connected chassis appears as disconnected in the configuration tree in MAX, select
Self-Test or Reset Chassis. If successful, the chassis icon changes to blue.
Figure 1-6. MAX Icons and States
For additional troubleshooting resources for the cDAQ chassis, refer to the Finding a Network
DAQ Device in MAX topic in the Measurement & Automation Explorer Help for NI-DAQmx.